Output intensity of LED light curing units in dental offices of Piura, Peru

Authors

Keywords:

dental office, visible light, electromagnetic waves, intensity, dental equipment, useful life.

Abstract

Introduction: Light curing lamps that use light-emitting diodes (LED) are used in dentistry for the polymeric conversion of dental restorative materials. It has been reported that inadequate light intensity in the lamp would not ensure the appropriate polymerization of restorative materials.

Objective: Determine the output intensity of LED light curing units used in dental offices of the city of Piura, Peru, in the year 2020.

Methods: An observational descriptive study was conducted. Measurements were taken of the light output of 70 LED light curing lamps using a radiometer with a wavelength of 400-500 nm and a light intensity measurement capacity of up to 3 500 mw/cm2. Intensity below 400 mw/cm2 was recorded as low, from 400 to 800 mw/cm2 as medium, from 800 a 1 200 mw/cm2 as high and above 1 200 mw/cm2 as very high.

Results: Intensity was medium in 48.5% of the lamps analyzed, high in 22.86%, low in 15.71% and very high in 12.86%. A lower frequency of lamps with greater clinical use was reported.

Conclusions: The LED light curing lamps used in dental offices of the province of Piura during the year 2020 emit an average output intensity of 778.14 mW/cm2, which corresponds to medium intensity.

Downloads

Download data is not yet available.

Author Biographies

Crista Gadith Palacios Rivas, Universidad César Vallejo, Facultad de Ciencias de la Salud, Escuela Profesional de Estomatología. Piura, Perú.

Cirujano Dentista. Escuela de Estomatología. Universidad César Vallejo. Piura, Perú.

Dora Denisse Cruz Flores, Universidad César Vallejo, Facultad de Ciencias de la Salud, Escuela Profesional de Estomatología. Piura, Perú.

Cirujano Dentista. Mg. en estomatología. Doctorando en Estomatología. Responsable de la Unidad de Investigación Escuela de Estomatología. Universidad César Vallejo. Piura, Perú.

Carmen Teresa Ibáñez Sevilla, Universidad Nacional de Trujillo, Escuela de Posgrado, Sección de Ciencias Médicas. Trujillo, Perú.

Cirujando Dentista. Especialista en Endodoncia. Magister en estomatología. Doctorando en Estomatología. Escuela de Posgrado.Seccción de Ciencias Médicas. Universidad Nacional de Trujillo.

MIGUEL ANGEL RUIZ-BARRUETO, Universidad César Vallejo, Facultad de Ciencias de la Salud, Escuela Profesional de Estomatología. Piura, Perú.

Biólogo Microbiólogo. Maestro en Ciencias con mención en Microbiología clínica. Doctor en Ciencias Biomédicas. Investgador Renacyt (Código Renacyt: P0015210). Docente de la Cátedra de Microbiología general y Estomatológica y de Proyecto y Desarrollo de tesis en Escuela de Estomatología. Universidad César Vallejo, Piura. Perú.

References

Abdulhaq A, Ahmad A, Hussein A. Effect of Contamination, Damage and Barriers on the Light Output of Light-Curing Units. Open Dent J [Internet]. 2019 [consulta: 30 julio 2019]; 13:196-202. Disponible en: https://www.researchgate.net/publication/334488048_Effect_of_Contamination_Damage_and_Barriers_on_the_Light_Output_of_Light-Curing_Units

Málaga J. Técnicas De Fotopolimerización Complementaria en La Evaluación De La Microdureza en Una Resina Compuesta De Nanopartículas. Revista Kiru [Internet]. Enero 2016 [consulta: 22 julio 2020];13(1):51–9. Disponible en: http://search.ebscohost.com/login.aspx?direct=true&db=a9h&AN=133435508&lang=es&site=ehost-live

Palin WM, Leprince JG, Hadis MA. Shining a light on high volume photocurable materials. Dent Mater [Internet] Marzo 2018 [consulta: 22 julio 2020];34(5):695-710. Disponible en: https://www.researchgate.net/publication/323840504_Shining_a_light_on_high_volume_photocurable_materials

Orozco R, Álvarez C, Guerrero J. Fotopolimerización de resinas compuestas a través de diversos espesores de tejido dental. Rev Odont Mex [internet]. Octubre 2015 [consulta: 12 octubre 2020]; 19(4): 222-227. Disponible en: https://doi.org/10.1016/j.rodmex.2015.10.002

Rueggeberg F, Giannini M, Galvao C, Precio R. Light curing in dentistry and clinical implications: a literature review. Braz. Oral Res [Internet]. 28 agosto 2017 [consulta: 28 Agosto 2020]; 31(61):64-91. Disponible en: http://dx.doi.org/10.1590/1807-3107bor-2017.vol31.0061

Soares C, Faria A, Rodrigues M, Fernandes A, Pfeifer C, Tantbirojn D, et al. Polymerization shrinkage stress of composite resins and resin cements what do we need to know. Braz. Oral Res [internet]. 28 agosto 2017 [consulta: 29 mayo 2020]; 31(suppl):e62 Disponible en https://doi.org/10.1590/1807-3107bor-2017.vol31.0062

Armellin E, Bovesecchi G, Coppa P, Pasquantonio G, Cerroni L. LED curing lights and temperature changes in different tooth sites. Biomed Res Int [Internet]. 18 abril 2016 [consulta: 08 marzo 2020]: 1-11. Disponible en: https://doi.org/10.1155/2016/1894672

Lamthong W, Yodyadthai K, Senawongse P. Output intensity of LED light curing units over a 4-year period of clinical use. M Dent J [Internet].2 agosto 2019 [consulta: 11 agosto 2020]; 39(2):91-101. Disponible en: https://dt.mahidol.ac.th/th/wp-content/uploads/2019/09/11-Mahidol-Dental-Journal-39-2.pdf

Silva R, Moura C, Miatto R, Gomes C, Pelissier B. Change in irradiance and energy density in relation to different curing distances. Braz Oral Res. [Internet]. 19 mayo 2015. [consulta: 15 marzo 2020]; 29(1):1-7. Disponible en https://www.scielo.br/scielo.php?script=sci_arttext&pid=S1806-83242015000100257&lng=en&nrm=iso&tlng=en DOI: 10.1590 / 1807-3107BOR-2015.vol29.0060

Bansal R, Bansal M, Walia S, Bansal L, Singh K, Aggarwal R. Assessment of efficacy and maintenance of light- curing units in dental offices across Punjab: A clinical survey. Indian J Dent Sci [Internet]. Enero 2019 [consulta: 10 abril 2020]; 11:42-5. Disponible en: http://www.ijds.in/text.asp?2019/11/1/42/252513

Mahmood M, Faraj B. Evaluation of light curing units and dentists knowledge about photo polymerization techniques in Sulaimani governmental dental clinics. Sulaimani Dent J [Internet]. 1 junio 2019 [consulta: 15 abril 2020]; 6(1):33-39. Disponible en: https://www.researchgate.net/publication/334631983_Evaluation_of_light_curing_units_and_Dentists'_knowledge_about_photo_polymerization_techniques_in_Sulaimani_governmental_dental_clinics

Cornejo J, Palamara M. Evaluation of intensity standards of tungsten- halogen and led curing units. J Oral Res [Internet]. 25 mayo 2019 [consulta: 15 abril 2020]; 8(2):104-107. Disponible en: http://www.joralres.com/index.php/JOR/article/view/joralres.2019.017

Ribeiro R, Carvalho F, Lima I, López A, Mezzalir H. Evaluación de la intensidad de luz y mantenimiento de los dispositivos. Fotopolimerizadores utilizados en las clínicas dentales en la ciudad de Recife- PE. Rev Odontol UNESP [Internet]. 12 diciembre 2016 [consulta: 10 agosto 2020]; 45(6):351-355. Disponible en: DOI: http://dx.doi.org/10.1590/1807-2577.06916

Muñoz R, Ampuero N. Efecto de las lámparas led en aclaramiento dental en la clínica odontológica UCSG semestre A-2017. Revista Conrado [Internet]. octubre 2017 [consulta: 12 febrero 2020]; 14(62):143-147. Disponible en: https://conrado.ucf.edu.cu/index.php/conrado/article/view/698/729

Marín L; Análisis de la ley de cuadrado inverso para medición práctica de iluminancia. Ingeniería [internet]. 26 mayo 2009 [consulta: 10 agosto 2020]; 19(1): 87-96. Disponible en: https://www.researchgate.net/publication/47276536_Analisis_de_la_ley_de_cuadrado_inverso_para_medicion_practica_de_iluminancia

Forghani N. The evaluation of the efficiency of LED light curing units used in private dental clinics. IJRRD [Internet].7 setiembre 2019 [consulta: 15 abril 2020]; 2(2):1-6. Disponible en https://www.journalijrrd.com/index.php/IJRRD/article/view/30100

Baharan R, Gosili A, Jaber M, Mahakhah A. Intensity output and effectiveness of light curing units in dental offices. J Clin Exp Dent [Internet]. 1 junio 2018 [consulta: 11 de abril 2020]; 10(6):555-60. Disponible en: https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6012498/

Lee Y, Abdul N, Isaqali M, Yusuf T, Syahrizal M. Evaluation of light- curing units used in dental clinics at a university in Malaysia. J Int Oral Health [Internet]. 28 agosto 2018 [citado 12 Julio 2020]; 10:206-9. Disponible en: https://www.jioh.org/text.asp?2018/10/4/206/240011

Hani M, Reem A, Hasanain F. Efficiency of light curing units in a government dental school. J Oral Sci [Internet]. Enero 2018 [consulta: 26 de mayo 2019]; 1:142-146. Disponible en https://www.researchgate.net/publication/323983635_Efficiency_of_light_curing_units_in_a_government_dental_school

Madhusudhana K, Venkata T, Suneekumar C, Lavanya A. A clinical survey of the output Intensity of light curing units in dental offices across Nellore urban area. SRM J Res Dent Sci [Internet]. 19 mayo 2016 [consulta: 19 de mayo 2020]; 7:64-8. Disponible en: https://www.srmjrds.in/text.asp?2016/7/2/64/182657

Published

2022-04-14

How to Cite

1.
Palacios Rivas CG, Cruz Flores DD, Ibáñez Sevilla CT, RUIZ-BARRUETO MA. Output intensity of LED light curing units in dental offices of Piura, Peru. Rev Cubana Estomatol [Internet]. 2022 Apr. 14 [cited 2025 Feb. 23];59(2):e3767. Available from: https://revestomatologia.sld.cu/index.php/est/article/view/3767